MODEL 842
362497
ISOBOX MODEL 842
ISOBOX MODEL 842
0.00

Product Description

The function of this reference unit differs from other cased systems in that the temperature of the metal block in which the thermocouple and copper leads are inserted, actually varies with ambient temperature. A separate output signal is produced which is proportional to the difference between the environmental temperature and the actual reference temperature. The output signal can be fed directly into a computer/data logger to give accurate compensation over a large ambient range. A thermal reservoir, heavily insulated, contains the reference junction probes. The reservoir temperature slowly follows the ambient temperature; an electrical compensation device is thermally integrated with the reservoir and thus senses the reservoir temperature. The device produces an output proportional to the difference between the reservoir temperature and the reference temperature (usually 0°C). This is the signal the computer/data logger uses to compensate for the temperature of the reference probes junctions.

The output signal can be in the form of a DC mV output, 4-20 mA or from a platinum resistance thermometer. Please discuss prior to order.

Units are housed in robust weatherproof enclosures to IP669 incorporating bottom gland plate. Fixing lugs for wall mounting are provided. Easy access to terminal rails and block assembly is via a lockable hinged front door.

Tech Spec

TECHNICAL FEATURE VALUE
Referencing Temp Effectively 0°C
Accuracy ±0.1°C per 10° ambient span
Long Term Stability ±0.05°C per 1000 Hours
Temperature Gradient ±0.1°C between Junctions
Stabilisation Time 10 minutes
Ambient Temperature -20°C to +70°C
Thermocouple 0 to 100 channels
Capacity Double junction referencing
Power 10 Watts typical 100-130 or 208-240 VAC 50/60Hz Low level D.C. Consumption 6VA typical Passive Option Available.
Dimensions Height 600mm Width 600mm Depth 300mm
Weight 40kg
Note Numerous special versions are available and can be supplied either in their existing form or modified to customers’ requirements.
How to order Model 842 – Isoboxes are normally uniquely specified for each order. Please discuss your exact requirements with us before ordering.

Downloads

  • Açıklama
    • Product Description

      The function of this reference unit differs from other cased systems in that the temperature of the metal block in which the thermocouple and copper leads are inserted, actually varies with ambient temperature. A separate output signal is produced which is proportional to the difference between the environmental temperature and the actual reference temperature. The output signal can be fed directly into a computer/data logger to give accurate compensation over a large ambient range. A thermal reservoir, heavily insulated, contains the reference junction probes. The reservoir temperature slowly follows the ambient temperature; an electrical compensation device is thermally integrated with the reservoir and thus senses the reservoir temperature. The device produces an output proportional to the difference between the reservoir temperature and the reference temperature (usually 0°C). This is the signal the computer/data logger uses to compensate for the temperature of the reference probes junctions.

      The output signal can be in the form of a DC mV output, 4-20 mA or from a platinum resistance thermometer. Please discuss prior to order.

      Units are housed in robust weatherproof enclosures to IP669 incorporating bottom gland plate. Fixing lugs for wall mounting are provided. Easy access to terminal rails and block assembly is via a lockable hinged front door.

      Tech Spec

      TECHNICAL FEATURE VALUE
      Referencing Temp Effectively 0°C
      Accuracy ±0.1°C per 10° ambient span
      Long Term Stability ±0.05°C per 1000 Hours
      Temperature Gradient ±0.1°C between Junctions
      Stabilisation Time 10 minutes
      Ambient Temperature -20°C to +70°C
      Thermocouple 0 to 100 channels
      Capacity Double junction referencing
      Power 10 Watts typical 100-130 or 208-240 VAC 50/60Hz Low level D.C. Consumption 6VA typical Passive Option Available.
      Dimensions Height 600mm Width 600mm Depth 300mm
      Weight 40kg
      Note Numerous special versions are available and can be supplied either in their existing form or modified to customers’ requirements.
      How to order Model 842 – Isoboxes are normally uniquely specified for each order. Please discuss your exact requirements with us before ordering.