Product Description
The Hyperion R Portable Blackbody Calibration Source allows for calibration of noncontact infrared thermometers over the temperature range -10°C to 80°C.
It is suitable for use as a primary radiation source for infrared thermometers from sub zero to 80°C.
Laboratory performance and low uncertainty calibrations are ensured by the combination by high emissivity and excellent temperature uniformity.
The digital temperature controller allows the block temperature to bet set to any value from -10°C to 80°C. Traceability of the radiance temperature is established by a separate, built-in temperature indicator and included platinum resistance thermometer.
A three point traceable calibration certificate is included. UKAS calibration of the resistance thermometer is available, as is radiometric calibration.
Uniformity of the block is ensured by using distributed thermoelectric heat pumps with the benefit of solid state, vibration free cooling.
Tech Spec
TECHNICAL FEATURE | VALUE |
---|---|
Model | 982 |
Temperature Range | -10°C to 80.0°C |
Emissivity | Greater than 0.995 |
Stability | ±0.1°C |
Display Resolution | 0.01°C |
Heating Time | 40 minutes to 80°C |
Cooling Time | 45 minutes to -10°C |
Aperture Diameter | 50mm |
Cavity Depth | 150mm |
PC Interface | Included |
Power | 200 Watts typical |
Voltage | 100-130 or 208-240 Vac |
Dimensions | Height 310mm Width 265mm Depth 200mm |
Weight | 10kg |
Options | Orifice Plates 10, 20, 30, 40mm (Restricts Cavity Aperture) – 812-01-06 Carrying Case 931-22-64 |
How to order | Model 982 Hyperion R: State Supply Voltage. Please state any special calibration requirements. |
Downloads
- Açıklama
Product Description
The Hyperion R Portable Blackbody Calibration Source allows for calibration of noncontact infrared thermometers over the temperature range -10°C to 80°C.
It is suitable for use as a primary radiation source for infrared thermometers from sub zero to 80°C.
Laboratory performance and low uncertainty calibrations are ensured by the combination by high emissivity and excellent temperature uniformity.
The digital temperature controller allows the block temperature to bet set to any value from -10°C to 80°C. Traceability of the radiance temperature is established by a separate, built-in temperature indicator and included platinum resistance thermometer.
A three point traceable calibration certificate is included. UKAS calibration of the resistance thermometer is available, as is radiometric calibration.
Uniformity of the block is ensured by using distributed thermoelectric heat pumps with the benefit of solid state, vibration free cooling.
Tech Spec
TECHNICAL FEATURE VALUE Model 982 Temperature Range -10°C to 80.0°C Emissivity Greater than 0.995 Stability ±0.1°C Display Resolution 0.01°C Heating Time 40 minutes to 80°C Cooling Time 45 minutes to -10°C Aperture Diameter 50mm Cavity Depth 150mm PC Interface Included Power 200 Watts typical Voltage 100-130 or 208-240 Vac Dimensions Height 310mm Width 265mm Depth 200mm Weight 10kg Options Orifice Plates 10, 20, 30, 40mm (Restricts Cavity Aperture) – 812-01-06 Carrying Case 931-22-64 How to order Model 982 Hyperion R: State Supply Voltage. Please state any special calibration requirements. Downloads