Product Description:
The 7050 multifunction process calibrator combines the essential functions of measurement and simulation of volts, millivolts, milliamps, and ohms with the direct readout simulation of thermocouples and RTD’s in °C or °F units.The instrument is microprocessor based and housed in a rugged case designed for use in the field. Input and Output functions are displayed simultaneously and the switch controls allow for easy and precise operation.
Five calibration points are instantaneously accessible through a preset percentage selector and all functions may be generated in a fully programmable ramp format.In addition to the standard calibration facilities the instrument may be used as a signal converter. Any electronic input signal can be converted and output as a proportional signal of 4-20mA, 0-5V or sq. root 4-20mA.
The 7050 is highly adaptable, IPT 68 and ITS 90 curves are accommodated and non-standard thermocouples and RTD’s may be programmed on request.
Technical Specification:
INPUT
VOLTAGE | ||
---|---|---|
RANGE | RESOLUTION | ACCURACY |
40V | 1mV | 0.03% |
4V | 100μV | 0.03% |
400mV | 10µV | 0.03% |
40mV | 1µV | 0.03% |
CURRENT | ||
---|---|---|
RANGE | RESOLUTION | ACCURACY |
400mA | 10μA | 0.03% |
40mA | 1μA | 0.03% |
4-20mA/0-100% | 0.01% | 0.06% |
RESISTANCE | ||
---|---|---|
RANGE | RESOLUTION | ACCURACY |
40000Ω | 1Ω | 0.03% |
4000Ω | 0.1Ω | 0.03% |
400Ω | 0.01Ω | 0.06% |
OUTPUT
VOLTAGE | ||
---|---|---|
RANGE | RESOLUTION | ACCURACY |
-2 to +10V | 1mV | 0.06% |
-100mV to +400mV | 10μV | 0.03% |
-10mV to +40mV | 1µV | 0.03% |
Output Impedance: < 10Ω
CURRENT | ||
---|---|---|
RANGE | RESOLUTION | ACCURACY |
0 to 20mA | 10μA | 0.03% |
4 to 20mA | 1μA | 0.03% |
TX Sim (4 to 20mA) | 0.01% | 0.03% |
Max Load Resistance: 900Ω at 20mA, Open circuit voltage 18- 27V
RESISTANCE | ||
---|---|---|
RANGE | RESOLUTION | ACCURACY |
0 to 400Ω | 0.01Ω | ± 0.03% |
Excitation Current 0.2 to 2mA
Other Specifications: Loop drive supply: Nominal 24V, current limit set to 30mA
Process Signal and Temperature Measurement & Simulation
TEMPERATURE | ||
---|---|---|
THERMOCOUPLE TYPE | TEMP RANGE °C | ACCURACY °C |
J | -210 to 0 0 to 710 710 to 1200 |
± 0.4 ± 0.2 ± 0.9 |
K | -270 to -240 -240 to -180 -180 to -50 -50 to 960 960 to 1370 |
± 3.0 ± 1.2 ± 0.5 ± 0.3 ± 1.7 |
T | -270 to -250 -250 to -180 -180 to -100 -100 to 400 |
± 3.0 ± 1.0 ± 0.5 ± 0.2 |
R | -50 to 0 0 to 70 70 to 400 400 to 1000 1000 to 1760 |
± 3.0 ± 2.0 ± 1.5 ± 1.0 ± 0.8 |
S | -50 to 0 0 to 100 100 to 450 450 to 1760 |
± 2.5 ± 1.8 ± 1.3 ± 0.9 |
B | 200 to 400 400 to 1000 1000 to 1820 |
± 5.0 ± 2.0 ± 1.0 |
N | 0 to 330 330 to 1100 1100 to 1300 |
± 0.4 ± 0.3 ± 1.5 |
E | -230 to 0 0 to 540 540 to 1000 |
± 0.5 ± 0.2 ± 1.5 |
RTD –
PT100 | |
---|---|
RANGE °C | ACCURACY °C |
-200 to 850 | ± 0.3 |
Thermocouple | Linearisation to BS 4937. ITS 90 or IPTS68 selectable. |
Pt100 | Linearisation to BS1904 (1984).DIN43760 (1980). Linearisation Acc: 0.1% of reading for Class A, 38.5Ω F.I |
Excitation current | 0.5mA (Input), 200μA to 2mA (Output). |
CJC | External (Pt100), Internal, Ice point. Accuracy +/-0.2°C at 23°C |
Ramp | Fully programmable continuous and stepped |
Steps | 5 Steps with fully adjustable zero |
Transmitter | Tx function 4-20mA |
Signal Converter | Any input to any output |
Dual Readout | Measure and source simultaneously displayed |
Stability | 100ppm per °C range 0 to 40°C |
Displays | LCD Dot Matrix, 1 line by 20 charx12mm, alphanumeric. |
Dimensions / Weight | W270 x H175 x D250mm, weight 4.5kg |
Optional Extras | Calibration Certificates traceable to NPL and UKAS |
- Açıklama
Product Description:
The 7050 multifunction process calibrator combines the essential functions of measurement and simulation of volts, millivolts, milliamps, and ohms with the direct readout simulation of thermocouples and RTD’s in °C or °F units.The instrument is microprocessor based and housed in a rugged case designed for use in the field. Input and Output functions are displayed simultaneously and the switch controls allow for easy and precise operation.
Five calibration points are instantaneously accessible through a preset percentage selector and all functions may be generated in a fully programmable ramp format.In addition to the standard calibration facilities the instrument may be used as a signal converter. Any electronic input signal can be converted and output as a proportional signal of 4-20mA, 0-5V or sq. root 4-20mA.
The 7050 is highly adaptable, IPT 68 and ITS 90 curves are accommodated and non-standard thermocouples and RTD’s may be programmed on request.
Technical Specification:
INPUT
VOLTAGE RANGE RESOLUTION ACCURACY 40V 1mV 0.03% 4V 100μV 0.03% 400mV 10µV 0.03% 40mV 1µV 0.03% CURRENT RANGE RESOLUTION ACCURACY 400mA 10μA 0.03% 40mA 1μA 0.03% 4-20mA/0-100% 0.01% 0.06% RESISTANCE RANGE RESOLUTION ACCURACY 40000Ω 1Ω 0.03% 4000Ω 0.1Ω 0.03% 400Ω 0.01Ω 0.06% OUTPUT
VOLTAGE RANGE RESOLUTION ACCURACY -2 to +10V 1mV 0.06% -100mV to +400mV 10μV 0.03% -10mV to +40mV 1µV 0.03% Output Impedance: < 10Ω
CURRENT RANGE RESOLUTION ACCURACY 0 to 20mA 10μA 0.03% 4 to 20mA 1μA 0.03% TX Sim (4 to 20mA) 0.01% 0.03% Max Load Resistance: 900Ω at 20mA, Open circuit voltage 18- 27V
RESISTANCE RANGE RESOLUTION ACCURACY 0 to 400Ω 0.01Ω ± 0.03% Excitation Current 0.2 to 2mA
Other Specifications: Loop drive supply: Nominal 24V, current limit set to 30mA
Process Signal and Temperature Measurement & Simulation
TEMPERATURE THERMOCOUPLE TYPE TEMP RANGE °C ACCURACY °C J -210 to 0
0 to 710
710 to 1200± 0.4
± 0.2
± 0.9K -270 to -240
-240 to -180
-180 to -50
-50 to 960
960 to 1370± 3.0
± 1.2
± 0.5
± 0.3
± 1.7T -270 to -250
-250 to -180
-180 to -100
-100 to 400± 3.0
± 1.0
± 0.5
± 0.2R -50 to 0
0 to 70
70 to 400
400 to 1000
1000 to 1760± 3.0
± 2.0
± 1.5
± 1.0
± 0.8S -50 to 0
0 to 100
100 to 450
450 to 1760± 2.5
± 1.8
± 1.3
± 0.9B 200 to 400
400 to 1000
1000 to 1820± 5.0
± 2.0
± 1.0N 0 to 330
330 to 1100
1100 to 1300± 0.4
± 0.3
± 1.5E -230 to 0
0 to 540
540 to 1000± 0.5
± 0.2
± 1.5RTD –
PT100 RANGE °C ACCURACY °C -200 to 850 ± 0.3 Thermocouple Linearisation to BS 4937. ITS 90 or IPTS68 selectable. Pt100 Linearisation to BS1904 (1984).DIN43760 (1980). Linearisation Acc: 0.1% of reading for Class A, 38.5Ω F.I Excitation current 0.5mA (Input), 200μA to 2mA (Output). CJC External (Pt100), Internal, Ice point. Accuracy +/-0.2°C at 23°C Ramp Fully programmable continuous and stepped Steps 5 Steps with fully adjustable zero Transmitter Tx function 4-20mA Signal Converter Any input to any output Dual Readout Measure and source simultaneously displayed Stability 100ppm per °C range 0 to 40°C Displays LCD Dot Matrix, 1 line by 20 charx12mm, alphanumeric. Dimensions / Weight W270 x H175 x D250mm, weight 4.5kg Optional Extras Calibration Certificates traceable to NPL and UKAS