Product Description:
High Performance Calibration
A precision multi-product calibrator designed to rapidly calibrate a wide range of test & measurement instruments. With over 11 standard functions the 5025C can accurately test and verify both traditional and new test equipment. It is ideal for companies requiring a calibrator that delivers high performance over wide conditions, with excellent capabilities, functionality, and features for ease of use.The 5025C multifunction calibrator is fitted with AC/DC voltage to 1050V, AC/DC current to 22A, digital frequency to 10MHz, decade resistance to 1GΩ, simulated variable resistance from 10Ω to 40kΩ, conductance, capacitance, thermocouple and PT100 simulation.
An optional enhanced performance pack enables advanced calibration for up to 6.5 digit multimeters. It provides full range variable resistance to 120MΩ, extended temperature functions, extended capacitance, inductance, additional precision high DCV ranges and increased ACV frequency bandwidth. Also available are oscilloscope and power calibration options.
External adaptors are available for clamp meter calibration (up to 1100A) and optical tachometer calibration. Also available is a power amplifier which increases the 5025C’s AC/DC voltage output drive capabilities to 100mA. This addition allows even the most demanding analogue instruments to be tested.
Wide Workload Coverage
The 5025C is designed for multifunction calibration of various instruments and devices; bench and handheld multimeters, frequency meters, ohm meters, AC/DC millivoltmeters, thermocouple and PT100 meters, clamp meters, and many other measurement devices. By adding internal options the 5025C can calibrate oscilloscopes, timer/counters, power meters and more.
For more information see the 5025 Series Calibration Capabilities Chart showing the DUT coverage of the 5025E and 5025C calibrators.
Simple Operation
Functions and ranges are easily accessed from the front panel. Increase and decrease keys per digit, are used to quickly set the output value. Deviation control then enables the user to finely adjust the output value as a percentage (+/- 9.999%). All this information is shown on a clear, easy to read LED display.
Virtual Control Software
The 5025C is supplied as standard with Time Electronics’ windows based Virtual Control Software. It allows quick and direct input of any value required. The remote SCPI command is displayed to help users develop custom applications.
EasyCal Calibration Software
Connect the 5025C to a PC/Laptop (via RS-232, USB or GPIB) installed with Time Electronics’ EasyCal software and automate the calibration process. This provides increased speed of calibration and consistency of results. Produce calibration certificates and reports to ISO 9001,ISO 17025, and other international quality standards.
Technical Specification:
FUNCTIONS | ||
---|---|---|
FUNCTION | RANGE / VALUES | BEST 1 YEAR SPECIFICATION |
Voltage DC | 0 to ± 1050V | ± 15ppm of setting |
Current DC | 0 to ± 22A 1100A with clamp meter adaptor (option 9780) |
± 60ppm of setting |
Voltage AC | 1mV to 1050V (20Hz to 100kHz sine-wave) | ± 200ppm of setting |
Current AC | 10μA to 22A (20Hz to 5kHz sine-wave) 100A with current transformer (option 9790) 1100A with clamp meter adaptor (option 9780) |
± 0.05% of setting |
Thermocouple Simulation | -210 to 1820°C. Type J, K, R, T, S, B, E, N | ± 0.05°C |
Digital Frequency/Period | 0.1Hz to 10MHz / 100ns to 10s | ± 20ppm of setting |
Conductance | 1s to 1ns (decade values) | ± 20ppm of setting |
Decade Resistance | 1Ω to 1GΩ (decade values) | ± 20ppm of setting |
Simulated Resistance | 10Ω to 40MΩ (variable) | ± 0.02% of setting |
PT100 Simulation | -180°C to 850°C | ± 0.2°C |
Capacitance | 1nF, 10nF, 20nF, 50nF, 100nF, 200nF, 500nF, 1μF | ± 0.25% of setting |
Option 9770: Oscilloscope
CALIBRATION | ||
---|---|---|
FUNCTION | RANGE / VALUES | BEST 1 YEAR SPECIFICATION |
Amplitude | 2mV to 200V and 1mV to 2V 50Ω (Square-wave or DC) | ± 0.05% |
Frequency/Period | 0.1Hz to 100MHz / 10ns to 10s (fixed values 1, 2, 5 sequence) | ± 20ppm |
Duty Cycle | 3 frequencies: 100Hz, 1kHz, 10kHz, settable from 0 to 100% | – |
Fast-Rise | < 400ps. Bandwidth checking up to 400MHz | – |
Option 9783: Frequency Reference For timer/counter calibration |
0.1Hz to 10MHz enhanced frequency accuracy reference | ± 0.1ppm |
Option 9769: 2.2GHz Sweep | 50MHz to 2.2GHz levelled sine-wave (0.5, 1, 1.5V pk-pk) | Amplitude ± 1%, Frequency ± 20ppm |
Option 9797: Power
CALIBRATION | ||
---|---|---|
FUNCTION | RANGE / VALUES | BEST 1 YEAR SPECIFICATION |
Power | 22A, 1050V, 23kW, 45 to 400Hz | ACV: 0.03%, DCV: 0.01%. ACI: 0.1%, DCI: 0.03% |
Phase/Power Factor | ± 90°/ 0.00 to 1.00PF | ± 0.3° |
Option 9701: Enhanced Performance Pack (added values/types shown in bold where
APPLICABLE) | ||
---|---|---|
FUNCTION | RANGE / VALUES | BEST 1 YEAR SPECIFICATION |
Full Range Resistance | 1Ω to 120MΩ (variable) | ± 100ppm of setting |
RTD Simulation | -180 to 850°C. Type PT100, PT200, PT500, PT1000 | ± 0.03°C |
Extended Thermocouple Simulation | -210 to 2315°C. Type J, K, R, T, S, B, E, N, C, L, U | ± 0.05°C |
Extended Capacitance | 1nF, 10nF, 20nF, 50nF, 100nF, 200nF, 500nF, 1μF, 10μF, 20μF, 50μF, 100μF | ± 0.2% of setting |
Extended AC Voltage Frequency | 1mV to 200mV/300kHz, 200mV to 2V/1MHz | ± 0.05% of setting |
Enhanced DC High Voltage | 20 to 200V and 100 to 1020V additional ranges | ± 15ppm |
Inductance | 1mH, 1.9mH, 5mH, 10mH, 19mH, 50mH, 100mH, 190mH, 500mH, 1H, 10H | ± 0.1% of setting |
For external options please see the Models/Options tab.
Warm up | 30 minutes to full accuracy |
Standard Interfaces | GPIB (IEEE-488), RS-232, USB |
Temp Performance | Operating: 5 to 45°C. Calibration: 15 to 28°C. Storage: -10 to 50°C |
Operating Humidity | < 80% non condensing |
Operating Altitude | 0 to 3km. Non operating: 3km to 12km |
Line Power | 100 to 230V AC 50/60 Hz. 200W maximum |
Dimensions | W455 x H155 x D480mm |
Weight | 16.5kg |
Country of Origin | UK |
Supplied With | User manual, RS-232 cable, and USB adaptor/cable |
For a full list of options please see the Models/Options tab.
For full specifications and further details please see the PDF Data Sheets in the Download tab.
- Açıklama
Product Description:
High Performance Calibration
A precision multi-product calibrator designed to rapidly calibrate a wide range of test & measurement instruments. With over 11 standard functions the 5025C can accurately test and verify both traditional and new test equipment. It is ideal for companies requiring a calibrator that delivers high performance over wide conditions, with excellent capabilities, functionality, and features for ease of use.The 5025C multifunction calibrator is fitted with AC/DC voltage to 1050V, AC/DC current to 22A, digital frequency to 10MHz, decade resistance to 1GΩ, simulated variable resistance from 10Ω to 40kΩ, conductance, capacitance, thermocouple and PT100 simulation.
An optional enhanced performance pack enables advanced calibration for up to 6.5 digit multimeters. It provides full range variable resistance to 120MΩ, extended temperature functions, extended capacitance, inductance, additional precision high DCV ranges and increased ACV frequency bandwidth. Also available are oscilloscope and power calibration options.
External adaptors are available for clamp meter calibration (up to 1100A) and optical tachometer calibration. Also available is a power amplifier which increases the 5025C’s AC/DC voltage output drive capabilities to 100mA. This addition allows even the most demanding analogue instruments to be tested.
Wide Workload Coverage
The 5025C is designed for multifunction calibration of various instruments and devices; bench and handheld multimeters, frequency meters, ohm meters, AC/DC millivoltmeters, thermocouple and PT100 meters, clamp meters, and many other measurement devices. By adding internal options the 5025C can calibrate oscilloscopes, timer/counters, power meters and more.For more information see the 5025 Series Calibration Capabilities Chart showing the DUT coverage of the 5025E and 5025C calibrators.
Simple Operation
Functions and ranges are easily accessed from the front panel. Increase and decrease keys per digit, are used to quickly set the output value. Deviation control then enables the user to finely adjust the output value as a percentage (+/- 9.999%). All this information is shown on a clear, easy to read LED display.Virtual Control Software
The 5025C is supplied as standard with Time Electronics’ windows based Virtual Control Software. It allows quick and direct input of any value required. The remote SCPI command is displayed to help users develop custom applications.EasyCal Calibration Software
Connect the 5025C to a PC/Laptop (via RS-232, USB or GPIB) installed with Time Electronics’ EasyCal software and automate the calibration process. This provides increased speed of calibration and consistency of results. Produce calibration certificates and reports to ISO 9001,ISO 17025, and other international quality standards.Technical Specification:
FUNCTIONS FUNCTION RANGE / VALUES BEST 1 YEAR SPECIFICATION Voltage DC 0 to ± 1050V ± 15ppm of setting Current DC 0 to ± 22A
1100A with clamp meter adaptor (option 9780)± 60ppm of setting Voltage AC 1mV to 1050V (20Hz to 100kHz sine-wave) ± 200ppm of setting Current AC 10μA to 22A (20Hz to 5kHz sine-wave)
100A with current transformer (option 9790)
1100A with clamp meter adaptor (option 9780)± 0.05% of setting Thermocouple Simulation -210 to 1820°C. Type J, K, R, T, S, B, E, N ± 0.05°C Digital Frequency/Period 0.1Hz to 10MHz / 100ns to 10s ± 20ppm of setting Conductance 1s to 1ns (decade values) ± 20ppm of setting Decade Resistance 1Ω to 1GΩ (decade values) ± 20ppm of setting Simulated Resistance 10Ω to 40MΩ (variable) ± 0.02% of setting PT100 Simulation -180°C to 850°C ± 0.2°C Capacitance 1nF, 10nF, 20nF, 50nF, 100nF, 200nF, 500nF, 1μF ± 0.25% of setting Option 9770: Oscilloscope
CALIBRATION FUNCTION RANGE / VALUES BEST 1 YEAR SPECIFICATION Amplitude 2mV to 200V and 1mV to 2V 50Ω (Square-wave or DC) ± 0.05% Frequency/Period 0.1Hz to 100MHz / 10ns to 10s (fixed values 1, 2, 5 sequence) ± 20ppm Duty Cycle 3 frequencies: 100Hz, 1kHz, 10kHz, settable from 0 to 100% – Fast-Rise < 400ps. Bandwidth checking up to 400MHz – Option 9783: Frequency Reference
For timer/counter calibration0.1Hz to 10MHz enhanced frequency accuracy reference ± 0.1ppm Option 9769: 2.2GHz Sweep 50MHz to 2.2GHz levelled sine-wave (0.5, 1, 1.5V pk-pk) Amplitude ± 1%, Frequency ± 20ppm Option 9797: Power
CALIBRATION FUNCTION RANGE / VALUES BEST 1 YEAR SPECIFICATION Power 22A, 1050V, 23kW, 45 to 400Hz ACV: 0.03%, DCV: 0.01%.
ACI: 0.1%, DCI: 0.03%Phase/Power Factor ± 90°/ 0.00 to 1.00PF ± 0.3° Option 9701: Enhanced Performance Pack (added values/types shown in bold where
APPLICABLE) FUNCTION RANGE / VALUES BEST 1 YEAR SPECIFICATION Full Range Resistance 1Ω to 120MΩ (variable) ± 100ppm of setting RTD Simulation -180 to 850°C. Type PT100, PT200, PT500, PT1000 ± 0.03°C Extended Thermocouple Simulation -210 to 2315°C. Type J, K, R, T, S, B, E, N, C, L, U ± 0.05°C Extended Capacitance 1nF, 10nF, 20nF, 50nF, 100nF, 200nF, 500nF, 1μF, 10μF, 20μF, 50μF, 100μF ± 0.2% of setting Extended AC Voltage Frequency 1mV to 200mV/300kHz, 200mV to 2V/1MHz ± 0.05% of setting Enhanced DC High Voltage 20 to 200V and 100 to 1020V additional ranges ± 15ppm Inductance 1mH, 1.9mH, 5mH, 10mH, 19mH, 50mH, 100mH, 190mH, 500mH, 1H, 10H ± 0.1% of setting For external options please see the Models/Options tab.
Warm up 30 minutes to full accuracy Standard Interfaces GPIB (IEEE-488), RS-232, USB Temp Performance Operating: 5 to 45°C. Calibration: 15 to 28°C. Storage: -10 to 50°C Operating Humidity < 80% non condensing Operating Altitude 0 to 3km. Non operating: 3km to 12km Line Power 100 to 230V AC 50/60 Hz. 200W maximum Dimensions W455 x H155 x D480mm Weight 16.5kg Country of Origin UK Supplied With User manual, RS-232 cable, and USB adaptor/cable For a full list of options please see the Models/Options tab.
For full specifications and further details please see the PDF Data Sheets in the Download tab.